ACP8820 + 2XACP8820X Optical Power Meter with Optical Head

ACP8820 + 2XACP8820X is an optical power meter with optical heads which offers superior performances in testing of DWDM components, AWG & PLC components, optical amplifiers, and other general purpose fiber optical test and measurement applications. It is designed especially for volume production line applications.

ACP8820 + 2XACP8820X Optical Power Meter is a high performance, small dimension, fast startup, affordable optical power meter family. It provides low power, high power, single channel and dual channel optical power meter modules options.  AC Photonics also can provide high channel count solutions up to 16 channels.

Features
  • High Performance
  • Quick Startup
  • Difference Power Range and Difference Channel Number Available
  • Small Dimension
  • Affordable Price
Applications
  • WDM, AWG, PLC Components Insertion Loss, Return Loss Test
  • Fiber Sensor, Fiber Cable Test
  • PMD and PDL Measurement
  • Fiber Optical, Telcom R and D Lab Test

Specifications

Parameter Unit Specifications
Model# ACP8820 + ACP88201 ACP8820 + 2XACP88201
Sensor Element Single Channel InGaAs Dual Channel InGaAs
Wavelength Range nm 850 ~ 1700
Power Range dBm +3 ~ -85
Application Fiber Type Standard SM and MM up to 62.5um core size
Uncertainty (accuracy) at Reference Condition nm ± 4% (1200 ~ 1610)
Relative Uncertainty (accuracy) at Reference Condition dB < 0.02 (Typical)
Linearity (power) dB ≤± 0.06 (1200nm ~ 1610nm, +0~ -60dBm)
Return Loss dB > 40
Operating Temperature °C 0 to +40
Storage Temperature °C - 30 to +80
Recalibration Period dB 2 Years
Package Dimensions (HxWxD) mm 200 x 105 x 250
Weight kg 6.5
Parameter Specifications
Model# ACP8820 + ACP88203 ACP8820 + 2XACP88203
Sensor Element Single Channel InGaAs Dual Channel InGaAs
Wavelength Range nm 850 ~ 1700
Power Range dBm +23 ~ -60
Application Fiber Type Standard SM and MM up to 62.5um core size
Uncertainty (accuracy) at Reference Condition nm ± 4% (1200 ~ 1610)
Relative Uncertainty (accuracy) at Reference Condition dB < 0.02 (Typical)
Linearity (power) dB ≤± 0.06 (1200nm ~ 1610nm, +0~ -60dBm)
Return Loss dB > 40
Operating Temperature °C 0 to +40
Storage Temperature °C - 30 to +80
Recalibration Period 2 Years
Package Dimensions (HxWxD) mm 200 x 105 x 250
Weight kg 6.5